Spo 2 Test; Nibp Tests - Mindray T1 Service Manual

Patient monitor
Hide thumbs Also See for T1:
Table of Contents

Advertisement

5.3.5 SpO
Test
2
Tool Required:
None.
Follow this procedure to perform the test:
1.
Connect SpO
sensor to the SpO
2
monitor.
2.
Apply the SpO
sensor to the ring finger of a healthy person.
2
3.
Check the Pleth wave and PR reading on the screen and make sure that the displayed SpO2 is within 95% and 100%. If
you are unable to get the SPO
4.
Remove the SpO
NOTE
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor. However, it can be used to
demonstrate that a particular pulse oximeter monitor reproduces a calibration curve that has been
independently demonstrated to fulfill a particular accuracy specification.

5.3.6 NIBP Tests

5.3.6.1 NIBP Accuracy Test
Tools required:
T-shape connector
Appropriate tubing
Squeeze bulb
Rigid Vessel with volume 500 ± 25 ml
Reference manometer (calibrated with accuracy equal to or better than 0.75 mmHg)
5-6
connector of the monitor. Set [Patient Cat.] to [Adu] and [PR Source] to SpO
2
between 95% and 100%, contact Mindray Technical Support.
2
sensor from the finger and make sure that an alarm of SpO
2
Sensor Off is triggered.
2
T1 Service Manual
on the
2