5.3 Performance Tests

Performance test are designed to ensure that measurement results are accurate. The following sections provide a list of
performance and accuracy tests and their recommended frequencies.

5.3.1 Performance Test Frequencies

Visual Inspection

ECG test
Resp performance test
SpO
test
2
NIBP test
Temp test
IBP performance test
CO
tests and
2
calibration
Analog output performance test
5.3.2 Visual Inspection
Inspect the equipment for obvious signs of damage. Follow these guidelines when inspecting the equipment:
Carefully inspect the case, display screen, buttons and knob for obvious signs of damage.
Inspect the power cord, and module accessories for obvious signs of damage
Inspect all external connections for loose connectors, bent pins or frayed cables.
Make sure that safety labels and data plates on the equipment are clearly legible.
5-4
Performance test
Verification
Pressure check
Leakage test
Leakage test
Performance test
Calibration
1. When first installed or reinstalled.
1. If the user suspects that the measurement is incorrect.
2. Following any repairs or replacement of relevant module.
3. Once a year for NIBP and CO
4. Once every two years for other parameter module performance tests.
If the user suspects that the analog output does not work.
tests.
2
T1 Service Manual