5.4.5.2 Battery Performance Test
1.
Turn off the monitor. Disconnect the monitor from the external power supply if DC adapter or T1 docking station is
connected.
2.
Install the battery in the monitor. Connect the external power supply and allow the battery to be charged
uninterruptedly till it is fully charged.
3.
Disconnect the external power supply. Remove the battery from the monitor. Keep the battery in room temperature for
two hours.
4.
Allow the monitor to run from the battery until the battery is completely depleted and the monitor automatically shuts
off.
The operating time of the batteries reflects their performance directly. If the operating time of a battery is noticeably shorter
than that stated in the specifications, replace the battery.

5.4.6 Network Print Test

Note
Use the recommended printers specified in the operator's manual (PN: 046-006734-00).
Tools required:
Hub and network cable
1
Connect the patient monitor and network printer to a HUB using common network cables as follows:
2
Set IP address as follows: Select [Main Menu]→ [Maintenance >>]→ [User Maintenance >>]→ enter the required
password→ [Network Setup >>]→ [Monitor Network Setup >>], set the IP address of the patient monitor in the
same network segment with that of the network printer. (See the instructions for use accompanying the printer)
3
Search for printer by selecting [Main Menu]→ [Print Setup >>]→ [Printer Setup >>]→ [Search Printer]. After a
while, the printer's model and IP address will appear in the box beside [Printer].