7.4.3.2 Removing the Power Board
Disconnect the cable from the socket on the power board, unscrew the four M3×6 screws indicated below and take out the
power board.
Unscrew the four
M3×6 screws

7.4.4 Disassembling the Top Housing Assembly

7.4.4.1 Removing the Locating Pin
Unscrew the two M3×6 screws, remove the locating pin pressing plate and take out the locating pin and locating pin spring.
Unscrew the two
M3×6 screws
7-16