7.3.3.2 Removing the Parameter Connector Panel Assembly
1.
Withdraw the analog out cable from the hole beside the fan on the main frame, and then disconnect the NIBP tubing.
Withdraw the analog out cable
from the hole beside the fan
2.
Unscrew the three M3×8 screws that secure the parameter connector assembly and remove the assembly from the
main frame.
7-8
Disconnect the NIBP tubing
T1 Service Manual