2.
Unscrew the two M3×8 screws securing the NIBP module retaining plate. Remove the NIBP module retaining plate.
3.
Disconnect the air tubing connecting the pump, the dump valve and the linear valve.
4.
Remove the pump and the valves. Before removing the pump and the valves, take the valve and pump cables from the
main frame to avoid damaging the cables.
T1 Service Manual
Unscrew the two M3×8 screws
Disconnect the air nozzle from the
pump and the valves
Dump and linear valves
Pump
7-11