C.7 HFA II - i To Third Party Programs Serial Data Transfer
Transfer of serial data from the HFA II - i to a third party program such as Peridata© is possible
only in the Serial II transfer mode.
Transmitting HFA II - i Tests to Third Party Programs
Ensure that the Serial II Transfer Disk is loaded by checking the destination window of the
Transfer Files selection on the Files Menu. If not, load the Serial II Transfer Disk as described
in Appendix section C.2.
From the HFA II - i Main Menu, select System Setup.
1)
Select the Save/Transmit option.
2)
Select RS-232 Options.
3)
4)
On the RS-232 Options screen, select the following settings:
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
Press Proceed. The Transmit/Save screen will then appear.
5)
Press the Main Menu icon in the upper right corner of the screen.
6)
From the Main Menu, select File Functions.
7)
Select Transfer Tests. A dialog box titled "Disk Options" will appear on screen.
8)
Select Hard Drive as the Source.
9)
Select HFA I Serial Cable as the Destination.
10)
Select the Directory Order (either Name or Date), then press Proceed.
11)
• If you chose Name:
To select all files on the drive, press Enter, and Select All. All files on the
disk will be marked with a "
• If you search files by patient name:
Enter the name - or just the first few characters - of the patient's last name whose test(s)
you wish to transfer. Select the test(s) you wish to transfer. All files selected will be
marked with a " " to indicate that they have been selected.
52235C1105
" to indicate that they have been selected.
Confidential and Proprietary
C - 11
Appendix C
Field Service Guide