Page 10
KNOWN ERRORS/ ISSUES - Application Software Rev. 3.6 - XML
•
The HFA2P network will not function if an active network connection is not available at boot time.
•
The FTP password entered using versions prior to version 3.6(0500), will get corrupted because the new
version expects an encrypted password. The FTP password needs to be re-entered with the new version.
•
Kinetic exams are not transferred.
•
User can only specify one server IP address for Data Export and TCP/IP transfers in HFA II - Serial format.
This release does not support specifying two servers independently.
RESTRICTIONS / LIMITATIONS
•
Current release does not support exporting of XML files for Kinetic exams.
•
Only single page prints can be exported in bitmap file format.
NOTE: This software cannot be ordered.
•
SOFTWARE FEATURES - Rev. 4.0 Application Software (P/N 57352)
Networking Features Standard (Optional for all models)
•
Archive patient data, test results and image files (tif format) to a file server.
•
Synchronize databases on two or more HFA2i perimeters.
•
Backup patient data and test results to a file server for safe external storage or off-line backup.
•
Restore patient data and test results from the network server to the HFA2i.
•
Export image files of printouts.
•
Must be licensed.
HFA-NET Pro Features (Optional for all models except 720i)
•
Import Worklists from certified EMR system.
•
Export patient data, image files (tif format) and test results to a certified EMR system.
•
Must be licensed.
Sita SWAP Features (Standard with 750i and 745i models)
•
Performs a SWAP (Short Wavelength Automated Perimetry) test using the SITA (Swedish Interactive Test
Algorithm).
•
Must be licensed.
GPA Features (Standard with 750i, 745i and 740i models).
•
Enhanced performance Glaucoma Progression Analysis.
•
Must be licensed.
Additional Features
•
Improved performance of Change Analysis Printout when analyzing 30-2 only printouts.
•
Reduction of name and exam association errors that may arise when saving tests.
•
Elimination of Serial Options Disk when performing a data transfer.
•
Elimination of Transfer Only choice on Save and Transmit Options
•
Improved readability of floppies created on an HFA II.
•
Increased name and ID field length from 24 characters to 60 characters.
FA2i-005J