KNOWN ERRORS/ ISSUES - Application Software Rev. 3.6 - XML
The HFA2P network will not function if an active network connection is not available at boot time.
The FTP password entered using versions prior to version 3.6(0500), will get corrupted because the new
version expects an encrypted password. The FTP password needs to be re-entered with the new version.
Kinetic exams are not transferred.
User can only specify one server IP address for Data Export and TCP/IP transfers in HFA II - Serial format.
This release does not support specifying two servers independently.
RESTRICTIONS / LIMITATIONS
Current release does not support exporting of XML files for Kinetic exams.
Only single page prints can be exported in bitmap file format.
NOTE: This software cannot be ordered.
SOFTWARE FEATURES - Rev. 4.0 Application Software (P/N 57352)
Networking Features Standard (Optional for all models)
Archive patient data, test results and image files (tif format) to a file server.
Synchronize databases on two or more HFA2i perimeters.
Backup patient data and test results to a file server for safe external storage or off-line backup.
Restore patient data and test results from the network server to the HFA2i.
Export image files of printouts.
Must be licensed.
HFA-NET Pro Features (Optional for all models except 720i)
Import Worklists from certified EMR system.
Export patient data, image files (tif format) and test results to a certified EMR system.
Must be licensed.
Sita SWAP Features (Standard with 750i and 745i models)
Performs a SWAP (Short Wavelength Automated Perimetry) test using the SITA (Swedish Interactive Test
Algorithm).
Must be licensed.
GPA Features (Standard with 750i, 745i and 740i models).
Enhanced performance Glaucoma Progression Analysis.
Must be licensed.
Additional Features
Improved performance of Change Analysis Printout when analyzing 30-2 only printouts.
Reduction of name and exam association errors that may arise when saving tests.
Elimination of Serial Options Disk when performing a data transfer.
Elimination of Transfer Only choice on Save and Transmit Options
Improved readability of floppies created on an HFA II.
Increased name and ID field length from 24 characters to 60 characters.