Instrument: Humphrey Field Analyzer II-i
Subject: System Validation Form - Prevail
Status: Mandatory - Domestic Only
Domestic Field Service:
A new System Validation Form has been released for the HFA II and HFA II-i. The System Validation
Form is required for all HFA II and HFA II-i field service calls. The values for the System Validation Form
are to be entered into the Prevail system when filling out your Field Service Report.
You are still required to perform the tasks specified in the System Checkout Checklist in Section 2 of the
Service Manual, but you will no longer need to submit the checklist with your Service Report. Values
obtained during the System Checkout will be entered into the System Validation Form on line.
The fields required on the System Validation Form are dependent on what type of Service Call, and
whether or not the instrument is equipped with B/Y hardware.
Service Call Types
Basic Service Call – Calls that do not require Intensity Verification or Calibration
Intensity Verification – Calls that require Intensity Verification only
Intensity Calibration – Calls that require Intensity Calibration and Verification
Attached is an example of the new System Validation Form.
International:
Continue to follow the current Checkout Procedures detailed in the Service Manual.
ÚÚÚ PROPRIETARY and CONFIDENTIAL ÚÚÚ
SERVICE
BULLETIN
No.: FA2i-026
Date:
7 Apr 04
Pages: 4