TABLE H - 1. CALIBRATION TOLERANCE RANGES
NVM DATA
Proj
Bowl
Blue
Yellow
Right Mode
Left Mode
Offset Fix
Focus
Shutter
Aperture
Color
CAMERA/GAZE
DATA
Illumination:
Reflex LED
Left LED
Right LED
Gain No Trial Lens
Gain Trial Lens
Scale
Field Service Guide
Humphrey Field Analyzer II - i
Min
Max
Voltage
40
127
10
127
40
127
10
127
Microsteps (X)
–1000
1000
–1000
1000
–1000
1000
Microsteps
–1000
1000
120
220
–1000
1000
–1000
1000
Relative Numbers
1
127
1
127
1
127
0
127
0
127
(not used)
(continued)
Confidential and Proprietary
H - 6
Min
Max
Intensity
30000
65534
1000
10000
60000
65534
10000
65534
Microsteps (Y)
–1000
1000
–1000
1000
–1000
1000
52235C1105