Hfa Ii To Hfa Ii - I Serial Data Transfer - Zeiss humphrey HFA II-i series Field Service Manual

Field analyzer
Table of Contents

Advertisement

Humphrey Field Analyzer II - i
C.4 HFA II To HFA II - i Serial Data Transfer
Transfer of serial data from the HFA II to HFA II - i is possible in both Serial I transfer mode
and Serial II transfer mode. However, you must have the HFA II - i and the HFA II set to the
Serial II transfer mode to ensure that all of the test information (SITA and gaze data) is passed
to the HFA II - i.
Transmitting HFA II Tests to HFA II - i
1)
Ensure that the Serial II Transfer Disk is loaded on the HFA II. Check the destination
window of the Transfer Files selection on the File Functions Menu. If not, load the Serial
II Transfer Disk as described in Appendix section C.2.
2)
Ensure that the data transfer cable is connected between the HFA II - i DATA
TRANSFER port and the HFA II SERIAL PORT 1.
On the HFA II
From the Main Menu, select System Setup.
3)
Select the Save/Transmit option.
4)
Select RS-232 Options.
5)
6)
On the RS-232 Options screen, select the following settings:
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
Press Proceed. The Transmit/Save screen will then appear.
7)
Press the Main Menu icon in the upper right corner of the screen.
8)
From the Main Menu, select File Functions.
9)
Select Transfer Tests. A dialog box titled "Disk Options" will appear on screen.
10)
Select Hard Drive as the Source.
11)
Select HFA II Serial Cable as the Destination.
12)
Select the Directory Order (either Name or Date), then press Proceed.
13)
• If you chose Name:
To select all files on the drive, press Enter, and Select All. All files on the disk
will be marked with a " " to indicate that they have been selected.
52235C1105
Confidential and Proprietary
C - 5
Appendix C
Field Service Guide