Appendix C
• If you search files by patient name:
Enter the name - or just the first few characters - of the patient's last name whose test(s)
you wish to transfer. Select the test(s) you wish to transfer. All files selected will be
marked with a " " to indicate that they have been selected.
• If you chose Date:
The file directory will be displayed in reverse chronological order. Enter a specific date
(mm-dd-yyyy), then choose Enter. The file directory will display tests taken on the
date you selected. Select the test(s) you wish to transfer. All files selected will be marked
with a " " to indicate that they have been selected.
Press Proceed. A pop up dialog box will appear advising you of the number of tests
14)
selected to be transferred. STOP HERE! Do Not choose Yes to start the transfer
process.
On the HFA II - i:
From the Main Menu, select System Setup.
15)
Select the Save/Transmit option.
16)
Select Serial Receive option.
17)
On the RS Receive Setup options screen, select the following settings:
18)
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
Press Proceed. The Transmit/Save screen will then appear. Press Proceed.
19)
The System Setup menu will appear.
20)
From the System Setup, select File Functions.
21)
Select Transfer Tests.
22)
Select HFA Serial Cable as the Source.
23)
Choose Hard or Floppy as the Destination, then Proceed.
24)
On the HFA II
To proceed, select YES.
25)
26)
After all tests are transmitted, a pop up dialog box will appear advising you of the
number of tests that were successfully transferred.
Field Service Guide
Humphrey Field Analyzer II - i
Confidential and Proprietary
C - 6