Specifications - Zeiss humphrey HFA II-i series Field Service Manual

Field analyzer
Table of Contents

Advertisement

General Information
Uninterruptible Power Supply (UPS)
C
Carl Zeiss Meditec does not sell a UPS (provides battery backup during power loss). However,
when selecting a UPS for use with the HFA II - i, note the following specifications and
recommendations:
The UPS must have a minimum 450 VA rating.
C
It must be dedicated to the HFA II - i. Do not connect the power table or other
C
devices into the UPS.
Recommended model: APC smart UPS 620.
C
Power backup during power loss will depend on the type of UPS selected. Consult a local
supplier for details.
HP LaserJet Printer
C
Carl Zeiss Meditec does not sell HP LaserJet Printers. Refer to Appendix D, Approved Parallel
Printers for details on using the HP LaserJet.
1.9

Specifications

HFA II - i Field Analyzer
Physical Characteristics
Dimensions . . . . . . . . . . . . . . . . . . 23" H x 22.25" W x 19.25" D
Weight . . . . . . . . . . . . . . . . . . . . . 87.6 lbs (36.3 kg)
Recommended Room Size . . . . . . 106" x 64" (269.2 x 162.6 cm)
Environmental
Operating Temp (ambient) . . . . . . 41 to 95E F
Humidity (ambient) . . . . . . . . . . . . 20 – 85%
Heat Generated . . . . . . . . . . . . . . 1,570 BTU/Hr (460 W)
Electrical
Input Power . . . . . . . . . . . . . . . . . 100 – 120 V @ 4 A, 50-60 Hz
Field Service Guide
Humphrey Field Analyzer II - i
(58.4 x 56.5 x 48.9 cm)
(Ventilation recommended)
230 V @ 2 A, 50 Hz
Confidential and Proprietary
1 - 22
52235C1105