Appendix M. Loading Application Software - Zeiss humphrey HFA II-i series Field Service Manual

Field analyzer
Table of Contents

Advertisement

Humphrey Field Analyzer II - i

Appendix M. Loading Application Software

Customer-installable software updates are supplied to the customer on floppy disks.
Currently, the software consists of four diskettes. These four diskettes are referred to as
the application system software, or simply the system software.
The system software consists of several other different software revisions, all of which
(except the installation software) get loaded into the instrument during the software
installation process. The system software consists of the following:
Installation software
C
Operating system software — loads onto the hard disk drive
C
Language software — loads onto the hard disk drive
C
CPU (boot) software — loads onto the hard disk drive
C
Motor Driver software — loads into the flash IC on the Motor Driver PCB
C
To view the revision levels of each software, refer to Appendix J, System Screens and
Logs.
In service situations where the hard disk is being replaced, or the service engineer
suspects possible corruption of the system software on the hard disk, the system software
can be reloaded onto the hard disk using the latest system software floppy disks.
Loading System Software:
1)
Switch the instrument power to OFF .
2)
Insert the 3.5" HFA II - i Series System Software floppy diskette labeled #1 into the
floppy disk drive. The current software requires four diskettes. Future software
versions may require more.
Switch the instrument power ON and you will soon see a VX Ld 1.2 .... display on
3)
the monitor as the operating system is loaded from the disk to the CPU memory.
When complete, you will see the following display:
HFA II - i Installer - Rev C1.1
0. Format the hard disk
1. Start Installation
4)
Select 1 to start the hard disk initialization. A Copying Files prompt will be seen
on the screen.
5)
When the disk files are loaded you will see a Please Insert Disk # 2 and press
any key dialogue displayed.
6)
Remove disk # 1 and insert disk # 2. Press any key on the keyboard.
52235C1105
Confidential and Proprietary
M - 1
Appendix M
Field Service Guide