G.1 Initializing the Cal/Config Data
This procedure is required whenever a new hard drive has been installed, before calibration
is performed.
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
From the DST Main Menu, select Configuration. The Configuration Menu
2)
appears.
Select Initialize the Configuration. A warning message appears informing
3)
you that all configuration and calibration data will be destroyed.
4)
Select OK. A test log screen appears confirming whether or not you want to continue.
Select Initialize to erase all contents of the Cal/Config Data. (Select Quit to
return to the Configuration Menu.)
5)
"Initialization successful" message will appear.
NOTE: If you initialize the CAL/Config Data, you must also perform the procedures for
Setting Serial Number, Setting the Model/Hardware Options, and Setting the Software
Options described below. Then recalibrate the instrument (see Table 3-1).
G.2 Setting Serial Number
This procedure is required any time the hard drive is replaced or the Cal/Config Data is
initialized. During this procedure you will be instructed to type in the serial number of the
instrument. The serial number can always be found on the label located on the rear cover
assembly.
NOTE: The instrument's serial number NEVER changes throughout the life of the
instrument. Instrument serial numbers are critical for tracking instrument reliability,
performance, and customer satisfaction.
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
From the DST Main Menu select Configuration. The Configuration Menu appears.
2)
Select Set Serial Number, then select Edit. A ten-key pad appears. Type in the
3)
four-digit model of the instrument. For example, 0750, then select [Enter].
4)
A ten-key pad appears. Type in the six-digit serial number of the instrument. For
example, 001000, then select [Enter].