Zeiss humphrey HFA II-i series Field Service Manual page 84

Field analyzer
Table of Contents

Advertisement

Parts Removal/Replacement
Replacement Notes:
Check/adjust the horizontal (X-axis) edge detector on the turret before reinstalling
P
the projection assembly into the instrument.
Adjust the belt as necessary (see Section 4).
P
When the three turret screws have been tightened, secure each one with a small
P
dab of Loctite® No. 222 (see Appendix A for part number).
When reassembled, manually rotate the turret to check for smooth movement
P
throughout its normal range.
Follow-up Checks/Adjustments:
Clean the optics. See Table 3-1.
Field Service Guide
Humphrey Field Analyzer II - i
Confidential and Proprietary
3 - 38
52235C1105