Color - Zeiss humphrey HFA II-i series Field Service Manual

Field analyzer
Table of Contents

Advertisement

Humphrey Field Analyzer II - i
5)
Make a penciled cross on a small piece of self-adhesive note paper and stick it into the
bowl such that the penciled cross is in the center of the III spot.
6)
Select aperture I; position the spot on the penciled cross using the STEP icons (UP
= left, DOWN = right); then select STORE.
7)
Repeat step 6 with all the other apertures. (II,IV,V)
4.8.2.4

Color

Purpose
This procedure is designed to ensure that all holes in the color wheel are properly centered
relative to the light path.
Procedure
1)
If the rear cover is on the instrument, remove the projection lamp access cover.
From the Mechanical Calibration menu, select COLOR, and note that the Clear 1 color
2)
wheel filter is initially selected.
3)
Observe the position of the hole in the color wheel in relation to the spot of light
striking the wheel.
52235C1105
FIGURE 4-5. Aperture Wheel
Confidential and Proprietary
4 - 15
Adjustment / Calibration
Field Service Guide

Hide quick links: