3.11 CRT PCB Removal

1)
Remove the operator panel assembly (3.4).
2)
Carefully discharge the CRT high voltage to ground and disconnect the high voltage
lead from the CRT.
3)
Disconnect:
the yoke connector from the CRT PCB.
P
the connector on the end of the CRT neck.
P
4)
Release and remove the CRT PCB from its four plastic standoffs.
Replacement Notes:
The CRT anode lead must be routed over the top of the CRT neck. If the lead passes
P
under the neck of the CRT, it lays next to the bowl and the electrostatic field around
the lead will attract dust to the inside of the bowl.
Follow-up Checks/Adjustments:
See Table 3-1.