Working with the Display

To scale measurement data

1
Select the active trace(s) you want to scale. (See "To select the active trace," the first
task in this section.)
2
For continuous autoscaling, press [
or
For single autoscaling, press [
press [
Autoscaling provides a convenient starting point when you are scaling measurement
data. It displays all of the data within the bounds of the active trace box. From this
starting point, you can easily reposition and expand the trace by modifying the values
of other scaling softkeys.
Matches Scale Menu scales the
active trace to any other trace
22
AUTOSCALE ON OFF
Use [
CENTER REFERENCE
reference value you specify. You can also anchor the reference to the top of the
trace box ([
TOP REFERENCE
([
BOTTOM REFERENCE
Use [
] to expand or compress the trace in relation to the Y-axis. You
Y PER DIV
enter a value that defines the height of each vertical division.
Use [
INP RANGE TRACKING
the input channel supplying the measurement data.
To learn more about the scaling softkeys, use online help.
] [
Scale
AUTOSCALE ON OFF
] [
Scale
AUTOSCALE ON OFF
] to highlight OFF.
] to expand or compress a trace around the center
]) or to the bottom of the trace box
]).
] to expand or compress a trace in relation to the level of
Y Per Dir defines the increment
between Y-axis divisions
Agilent 35670A Quick Start Guide
] to highlight ON.
] to highlight ON, then