Agilent 35670A Quick Start Guide
4
Specify the triggering parameters.
Press [
] [
Trigger
SOURCE TRIGGER
5
Configure the display.
Press [
], then press[
Scale
Press [
] [
Disp Format
6
Measure the DUT.
Press [
].
Start
A triggered measurement and the uniform window selection produces a much more
accurate response at the low end of the frequency spectrum. The periodic chirp
provides more energy than random noise. It is a self-windowing signal that provides
exactly the same stimulus from one measurement average to the next. Consequently,
averaging is not required.
The analyzer's source impedance is less than 5 ohms. If your DUT needs a specific
source impedance, insert the appropriate resistor in series with the source output.
To display the measurement data on a logarithmic scale
press [
] then press [
Trace Coord
].
] to highlight ON.
AUTOSCALE ON OFF
] [
].
MORE
BODE DIAGRAM
X-AXIS LIN LOG
Making Measurements
] to highlight LOG.
31