Making Measurements
To calibrate a microphone — Octave Analysis (Option 1D1)
Octave measurements are available only with those Agilent 35670A analyzers that are
equipped with Option 1D1. Signals can be analyzed at frequencies up to 40 kHz in the
single-channel mode. This task measures the sound pressure level of the calibrator.
1
Initialize the analyzer.
Press [
Press [
2
Connect the microphone input to the analyzer's Channel 1 input as shown in the
illustration. Attach a calibrator to the microphone.
3
Configure the display.
Press [
4
Start the calibration.
Turn on the calibrator.
Press [
Calibrator
Laboratory
Microphone
34
] [
].
Preset
DO PRESET
].
OCTAVE ANALYSIS
], then press[
Scale
AUTOSCALE ON OFF
].
Start
Agilent 35670A Quick Start Guide
] to highlight ON.