Making Measurements
To measure frequency response — FFT
A device's frequency response can be measured using the high-speed FFT analysis.
The dynamic range is typically 80 dB. For better dynamic range, use the optional
Swept Sine instrument mode.
1
Initialize the analyzer.
Press [
].
Preset
[
].
DO P RESET
Press [
] [
Meas Data
FREQ RESP 2/1
2
Connect the device-under-test (DUT) as shown in the illustration.
3
Specify the measurement parameters.
Press [
] [
] <number> <unit>, then press [
Freq
START
Press [
], [
Source
PERIODIC CHIRP
Press [
] <number> <unit>, then press [
LEVEL
Press [
] [
Window
UNIFORM
Source
30
].
].
SOURCE ON OFF
].