Resistance Verification - Keithley DMM7510 Calibration Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 2: Performance verification
Range
Nominal calibrator
value
(Ω)
100
0
100.0000
1000
0
1000.000
1.0E+04
0
1900.000

Resistance verification

To check the resistance function, you will:
Use shielded, Teflon-insulated or equivalent cables in a 4-wire configuration
Characterize the calibrator 1 Ω, 10 Ω, and 909 MΩ nominal values with external reference DMM
Verify that the displayed readings fall within specified limits
For the 1 Ω, 10 Ω, and 1 GΩ ranges, verify accuracy from reference DMM readings
For the 100 Ω to 100 MΩ ranges, verify accuracy from actual calibrator source values
Characterize the calibrator and 909 MΩ resistor
If not already done, complete the procedure for characterizing the calibrator for dry circuit verification.
Characterize the calibrator for 1 Ω and 10 Ω
Refer to
To characterize the resistor:
1. Make connections between the resistor and the reference DMM as shown in the figure below.
2-40
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Actual calibrator
value
(Ω)
9.9996340E01
9.9998410E+02
1.8999910E+03
Figure 4: 909 MΩ resistor characterization
Lower limit
(Ω)
-8.00E-03
9.997934E+01
-8.00E-02
9.997241E+02
-8.00E-01
1.898583E+03
(on page 2-38).
DMM7510-905-01 Rev. A / April 2015
Upper limit
(Ω)
8.00E-03
1.000133E+02
8.00E-02
1.000244E+03
8.00E-01
1.901399E+03