Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
--AC voltage adjustment step 6: 10 V at 1.0 kHz
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear()
triangle.
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #6 10V, 1KHz")
cal.adjust.ac(6)
code,message,severity = eventlog.next()
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #6 10V, 1KHz , Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')
--AC voltage adjustment step 7: 10 V at 50 kHz
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear()
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #7 10V, 50KHz")
cal.adjust.ac(7)
code,message,severity = eventlog.next()
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #7 10V, 50KHz, Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')
DMM7510-905-01 Rev. A / April 2015
-- Clear the error event log and Home screen error
-- 10V, 1KHz HI and LO Ft terminals
-- Set variables to print the error
-- Clear the error event log and Home screen error triangle.
-- 10V, 50KHz HI and LO Ft terminals.
-- Set variables to print the error