Cal.unlock() - Keithley DMM7510 Calibration Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual

cal.unlock()

This attribute unlocks calibration operations.
Type
TSP-Link accessible
Function
No
Usage
cal.unlock(password)
password
Details
Calibration data is locked during normal operation. To perform calibration, you must unlock
calibration.
The default password is KI000CAL. You can use cal.password to change the default.
An error is generated if this command is issued when calibration is already unlocked.
Example
cal.unlock("KI000CAL")
-- Perform operations to generate the calibration data
cal.save()
cal.lock()
Unlock the calibration for the instrument using the default password.
Save the calibration data.
Lock the calibration data.
Also see
cal.lock()
cal.password
DMM7510-905-01 Rev. A / April 2015
Affected by
A string containing the password to unlock calibration
(on page 4-16)
(on page 4-17)
Section 4: TSP command reference
Where saved
Default value
4-19