Section 4: TSP command reference

cal.adjust.dc()

This function starts the specified DC adjustment step.
Type
TSP-Link accessible
Function
Yes
Usage
cal.adjust.dc(step)
cal.adjust.dc(step, value)
step
value
Details
The descriptions of these steps are:
DC Step 0: Internal amplifier offset adjustment
DC Step 1: 4-wire zero for DCI 1 A to 3 A, DCV, Digitize DCV, 2W Ω, and 4W Ω
DC Step 2: Open Terminals:
DCI, zero 10 µA to 100 mA
DCV, 100 V and 1000 V full-scale
Resistance, 1 Ω to 1 GΩ full-scale
Digitize DCI and DCV full-scale
Waveform, analog trigger
Frequency, threshold level trigger
DC Step 3: 10 V, DCV 10 V range
DC Step 4: -10 V, DCV 10 V range
DC Step 5: 10 kΩ, Resistance, 4W 10 kΩ range
For detail on how to use cal.adjust.dc(), refer to
11).
This command generates an error if:
Calibration is locked
The step is out of sequence
The step does not complete successfully
The value that is passed is invalid for the step, out of range, or not needed
Also see
cal.adjust.ac()
cal.lock()
(on page 4-16)
cal.unlock()
(on page 4-19)
4-12
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Affected by
The DC adjustment step to start
The value for this adjustment step (if applicable)
(on page 4-8)
Where saved
Default value
DC voltage adjustment steps 0 to 5
DMM7510-905-01 Rev. A / April 2015
(on page 3-