Section 2: Performance verification
Verify the digitize voltage 100 mV range
Description
Perform rel offset
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
Verify the digitize voltage 1 V range
Description
Verify zero
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
Verify the digitize voltage 10 V range
Description
Verify zero
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
Verify the digitize voltage 100 V range
Description
Verify zero
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
Verify the digitize voltage 1000 V range
Description
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
2-18
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Input
Lower limit
0.0
n/a
0.1
9.99680E-02
0.05
4.99790E-02
-0.05
-5.00210E-02
-0.1
-1.00032E-01
Input
Lower limit
0
-7.50E-05
1
9.99805E-01
0.5
4.99865E-01
-0.5
-5.00135E-01
-1
-1.00020E+00
Input
Lower limit
0
-7.50E-04
10
9.99805E+00
5
4.99865E+00
-5
-5.00135E+00
-10
-1.00020E+01
Input
Lower limit
0
-7.50E-03
100
9.99805E+01
50
4.99865E+01
-50
-5.00135E+01
-100
-1.00020E+02
Input
Lower limit
1000
9.99805E+02
500
4.99865E+02
-500
-5.00135E+02
-1000
-1.00020E+03
Upper limit
n/a
1.00032E-01
5.00210E-02
-4.99790E-02
-9.99680E-02
Upper limit
7.50E-05
1.00020E+00
5.00135E-01
-4.99865E-01
-9.99805E-01
Upper limit
7.50E-04
1.00020E+01
5.00135E+00
-4.99865E+00
-9.99805E+00
Upper limit
7.50E-03
1.00020E+02
5.00135E+01
-4.99865E+01
-9.99805E+01
Upper limit
1.00020E+03
5.00135E+02
-4.99865E+02
-9.99805E+02
DMM7510-905-01 Rev. A / April 2015