Section 2: Performance verification
Verify the analog trigger voltage 100 mV range
Description
Measure or Digitize DC-coupled
50 % scale rising
Measure or Digitize DC-coupled
50 % scale falling
Digitize AC-coupled
50 % scale rising
Digitize AC-coupled
50 % scale falling
Verify the analog trigger voltage 1 V range
Description
Measure or Digitize DC-coupled
50 % scale rising
Measure or Digitize DC-coupled
50 % scale falling
Digitize AC-coupled
50 % scale rising
Digitize AC-coupled
50 % scale falling
Verify the analog trigger voltage 10 V range
Description
Measure or Digitize DC-coupled
50 % scale rising
Measure or Digitize DC-coupled
50 % scale falling
Digitize AC-coupled
50 % scale rising
Digitize AC-coupled
50 % scale falling
Verify the analog trigger voltage 100 V range
Description
Measure or Digitize DC-coupled
50 % scale rising
Measure or Digitize DC-coupled
50 % scale falling
Digitize AC-coupled
50 % scale rising
Digitize AC-coupled
50 % scale falling
2-24
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Verification point
Lower limit
0.05
0.0490
-0.05
-0.0510
0.05
0.0485
-0.05
-0.0515
Verification point
Lower limit
0.5
0.4900
-0.5
-0.5100
0.5
0.4850
-0.5
-0.5150
Verification point
Lower limit
5
4.900
-5
-5.100
5
4.850
-5
-5.150
Verification point
Lower limit
50
49.00
-50
-51.00
50
48.50
-50
-51.50
Upper limit
0.0510
-0.0490
0.0515
-0.0485
Upper limit
0.5100
-0.4900
0.5150
-0.4850
Upper limit
5.100
-4.900
5.150
-4.850
Upper limit
51.00
-49.00
51.50
-48.50
DMM7510-905-01 Rev. A / April 2015