Keithley DMM7510 Calibration Manual page 107

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 3: Calibration
--AC voltage adjustment step 8: 100 V at 1.0 kHz
--WARNING: The following calibration steps are intended for qualified service
personnel
-- only. Do not attempt these procedures unless you are qualified to do so. The
following
-- procedures require hazardous voltages, which could cause personal injury or
death if
-- contacted. Use appropriate safety precautions when working with hazardous
voltages.
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear()
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #8 100V, 1KHz")
cal.adjust.ac(8)
code,message,severity = eventlog.next()
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #8 100V, 1KHz, Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')
3-30
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
-- Clear the error event log and Home screen error triangle.
-- 100V, 1KHz HI and LO Ft terminals
-- Set variables to print the error
DMM7510-905-01 Rev. A / April 2015

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents