Keithley DMM7510 Calibration Manual page 66

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Verify AC current 1 mA range
Description
@ 40 Hz
@ 1 kHz
@ 5 kHz
Verify AC current 10 mA range
Description
@ 40 Hz
@ 1 kHz
@ 5 kHz
Verify AC current 100 mA range
Description
@ 40 Hz
@ 1 kHz
@ 5 kHz
Verify AC current 1 A range
Description
@ 40 Hz
@ 1 kHz
@ 5 kHz
Verify AC current 3 A range
Description
2 A @ 40 Hz
2 A @ 1 kHz
2 A @ 5 kHz
DMM7510-905-01 Rev. A / April 2015
Verification point Lower limit
0.001
9.989000E-04
0.001
9.989000E-04
0.001
9.988000E-04
Verification point Lower limit
0.010
9.989000E-03
0.010
9.989000E-03
0.010
9.988000E-03
Verification point Lower limit
0.1
9.989000E-02
0.1
9.989000E-02
0.1
9.988000E-02
Verification point Lower limit
1.000
9.976000E-01
1.000
9.976000E-01
1.000
9.908000E-01
Verification point Lower limit
2.000
1.994500E+00
2.000
1.994500E+00
2.000
1.980900E+00
Section 2: Performance verification
Upper limit
1.001100E-03
1.001100E-03
1.001200E-03
Upper limit
1.001100E-02
1.001100E-02
1.001200E-02
Upper limit
1.001100E-01
1.001100E-01
1.001200E-01
Upper limit
1.002400E+00
1.002400E+00
1.009200E+00
Upper limit
2.005500E+00
2.005500E+00
2.019100E+00
2-57