Simulated Rtd Temperature Verification - Keithley DMM7510 Calibration Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 2: Performance verification
2. Set the calibrator to 0 V and enable the output.
3. Allow 5 minutes of settling time.
4. On the Model DMM7510, press the FUNCTION key and select DC voltage.
5. Press the HOME key.
6. Select Range and select 100 mV.
7. Press MENU key. Under Measure, select Settings.
8. Set Input Impedance to Auto.
9. Record the voltage offset reading.
10. Subtract the DCV 100 mV reading from the calibrator source value.
11. On the Model DMM7510, press the FUNCTION key and select Temperature.
12. Press MENU key. Under Measure, select Settings.
13. Set Thermocouple to Type J.
14. Set the Simulated Temperature to 0°C.
15. Set the output DC Voltage to value shown in the following tables.
16. Subtract the DCV 100 mV calibrator and cable offset voltage.
Verify the simulated thermocouple Type J temperature
Table calibrator source values based on NIST Monograph 175, reference data 60, version 2.0.
Use the following values to verify the performance of the Model DMM7510. Actual values depend on
published specifications (see
Description
-190 °C
0 °C
750 °C

Simulated RTD temperature verification

Use the following to verify the performance of the Model DMM7510. Actual calibrator source values
will vary. RTD verification is based on the calibrator sourcing resistance and the Model DMM7510
conversion of the resistance measurement to calculated temperature based on the Callendar-Van
Dusen equation.
2-34
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Example reading limit calculation
Typical calibrator
Lower limit
source value
(V)
-7.659 E-03
-190.200 °C
0.0000 E-03
-0.200 °C
4.2281 E-02
749.800 °C
(on page 2-8)).
Upper limit
-189.800 °C
0.200 °C
750.200 °C
DMM7510-905-01 Rev. A / April 2015