Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
cal.adjust.internaltemp
This attribute returns the internal temperature of the instrument when calibration was last adjusted.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
temperature = cal.adjust.internaltemp
temperature
Example
print(cal.adjust.internaltemp)
Also see
cal.adjust.tempdiff
cal.adjust.rear.ac()
This function starts the specified rear AC adjustment step.
Type
TSP-Link accessible
Function
Yes
Usage
cal.adjust.rear.ac(step)
cal.adjust.rear.ac(step, value)
step
value
Details
The descriptions of these steps are:
AC Current Step 1, 0.5 A 1 kHz, AC current 10A range zero
AC Current Step 2, 2 A to 10 A 1 kHz, AC current 10A range full-scale
For detail on how to use cal.adjust.rear.ac(), see
For code examples, refer to:
•
Rear AC adjustment step 1: ACI 10 A Zero adjustment
•
Rear AC adjustment step 2: ACI 10 A Full-scale adjustment
DMM7510-905-01 Rev. A / April 2015
Affected by
Not applicable
The internal temperature in Celsius (°C)
(on page 4-15)
Affected by
The rear AC adjustment step to start
The value for this adjustment step (if applicable)
Where saved
Nonvolatile memory
Returns the internal temperature of the instrument when
calibration was last adjusted.
Example output:
34.4
Where saved
Rear terminal adjustments
(on page 3-8)
(on page 3-9)
Section 4: TSP command reference
Default value
Not applicable
Default value
(on page 3-6).
4-13