Dry Circuit Resistance Verification - Keithley DMM7510 Calibration Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
3. On the Model DMM7510, press the FUNCTION key and select Temperature.
4. Press the MENU key. Under Measure, select Settings.
5. Select Transducer.
6. Set the Type to either 4-wire RTD or 3-Wire RTD.
7. Set the RTD Type to PT100.
8. Press the HOME key.
9. On the calibrator, select 19 Ω source resistance
10. Select the OPER and EX SNS keys.
11. Record Model DMM7510 accuracies.
12. Refer to the table for PT100 accuracies.
Fluke 5720 and 5730 resistance source values vary and may require new resistance-to-temperature
target accuracy values.
13. Repeat for 100 Ω and 190 Ω source values.
Example
Nominal calibrator
value
(Ω)
19
100
190

Dry circuit resistance verification

To check the dry circuit resistance function, you will:
Use shielded, Teflon-insulated or equivalent cables in a 4-wire configuration
Characterize the calibrator 1 Ω and 10 Ω nominal values with the external reference DMM
Verify that the displayed readings fall within specified limits
For the 1 Ω and 10 Ω ranges, verify accuracy from the reference DMM readings
For the 100 Ω to 10 kΩ ranges, verify accuracy from actual calibrator source values
DMM7510-905-01 Rev. A / April 2015
PT100
Ro
alpha
beta
delta
A
B
C
Actual calibrator
Temperature
value
(°C)
(Ω)
18.999520
-198.8900
99.99707
-0.0075
189.99234
238.6775
1.000000E+02
3.850550E-03
1.086300E-01
1.499900E+00
3.908304E-03
-5.775440E-07
-4.182852E-12
4-wire RTD
±0.06 °C accuracy
(±Ω from actual
calibrator value)
0.0259
0.0235
0.0218
Section 2: Performance verification
3-wire RTD
±0.75 °C accuracy
(±Ω from actual
calibrator value)
0.3241
0.2932
0.2725
2-37