Section 3: Calibration
--Rear DC adjustment step 3: DCI 10 A full-scale adjustment
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear()
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step #3, 2A, DCA 10A range")
cal.adjust.rear.dc(3, 2.0)
code,message,severity = eventlog.next()
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step #3, Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')
--Rear AC adjustment step 1: ACI 10 A Zero adjustment
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear()
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step#1, 0.5A, ACI 10A Zero")
cal.adjust.rear.ac(1)
10A and Lo rear terminals.
code,message,severity = eventlog.next()
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step#1, ACI, Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')
3-22
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
-- Clear the error event log and Home screen error triangle.
-- Requires calibrator to source 2A.
-- Clear the error event log and Home Screen error triangle.
-- Requires Fluke Model 5700 to source 0.5A ac at 1KHz into
-- Set variables to print the error
-- Set variables to print the error