Advertisement

Section 3: Calibration
--AC current adjustment step 11: 100 uA at 1 kHz
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear()
triangle.
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #11, 100uA, 1KHz ")
cal.adjust.ac(11)
code,message,severity = eventlog.next()
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, 100uA, 1KHz , Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')
--AC current adjustment step 12: 1 mA at 1 kHz
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear()
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #12, 1mA, 1KHz")
cal.adjust.ac(12)
code,message,severity = eventlog.next()
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, 1mA, 1KHz , Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')
3-32
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
-- Clear the error event log and Home screen error
-- 100uA, 1KHz 3A and LO Ft terminals.
-- Clear the error event log and Home screen error triangle.
-- 1mA, 1KHz 3A and LO Ft terminals.
-- Set variables to print the error
-- Set variables to print the error
DMM7510-905-01 Rev. A / April 2015