Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
10 A verification connections
Ensure that the gauge of the cabling is sufficient to handle 10 A.
Verify the analog trigger current 10 A range
To verify the analog trigger current 10 A range:
1. Set the TERMINALS switch to REAR. Ensure that the orange R is displayed.
2. Set the calibrator and Model DMM7510 as shown in
Use the following values to verify the performance of the Model DMM7510.
Verify the analog trigger current 10 A range
Description
Measure or digitize
50 % scale rising
Measure or digitize
50 % scale falling
DMM7510-905-01 Rev. A / April 2015
Verification point
Lower limit
5.0000E+00
4.70E+00
-5.0000E+00
-5.30E+00
Section 2: Performance verification
10 A verification connections
Upper limit
5.30E+00
-4.70E+00
(on page 2-29).
2-29