Keithley DMM7510 Calibration Manual page 58

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Zero verify the Model DMM7510
1. On the calibrator, select the OPR/STBY key. Ensure that the front panel displays STANDBY.
2. Verify the Model DMM7510 Zero reading for each range.
Rel the system
1. On the calibrator, select the OPR/STBY key. Ensure that the front panel displays OPERATE.
2. Set the reference DMM to DC Current, 2 mA range.
3. Set the calibrator to source zero current.
4. Rel the Model DMM7510.
5. Zero Range the reference DMM for system offset.
6. Set the calibrator to source the current identified in the 1 mA verification, verifying the Model
DMM7510 accuracy from the calculated reference DMM current.
7. For 10 mA verification, set the reference DMM to the 20 mA range.
8. Repeat steps 1 to 6 for 10 mA verification.
Verify DC current 1 mA and 10 mA range
Range
1.0 e-03
1.0 e-02
DMM7510-905-01 Rev. A / April 2015
Nominal input
Typical measured
input (A)
standby
0.00E+00
8.00E-07
1.00E-03
9.9999100E-04
5.00E-04
4.9999740E-01
-5.00E-04
-4.9999540E-04
-1.00E-03
-9.9999040E-04
standby
0.00E+00
6.00E-06
1.00E-02
9.9999990E+00
5.00E-03
5.0000180E+00
-5.00E-03
-5.0000060E+00
-1.00E-02
-1.0000000E+01
Section 2: Performance verification
Lower limit (A)
Upper limit (A)
-9.00E-09
9.00E-09
REL
REL
9.9992200E-04
1.0000600E-03
4.9996739E-01
5.0002741E-01
-5.0003440E-04
-4.9995640E-04
-1.0000594E-03
-9.9992140E-04
-9.00E-08
9.00E-08
REL
REL
9.9993989E+00
1.0000599E+01
4.9997179E+00
5.0003181E+00
-5.0003061E+00
-4.9997059E+00
-1.0000600E+01
-9.9993999E+00
2-49

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents