Keithley DMM7510 Calibration Manual page 88

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
To adjust the Model DMM7510 using the front terminals:
1. Set the TERMINALS switch to FRONT. Ensure that the green F is displayed.
2. Install the Keithley Model 8610 or 8620 shorting plug on the front terminals of the Model
DMM7510, as shown in the figure below.
3. Allow to settle for 5 minutes.
4. Perform the cal.adjust.dc steps 0 through 5, cal.adjust.ac, Voltage, steps 1 through 10,
cal.adjust.ac, Current, step 11 through 16, cal.adjust.ac, coupling amplitude adjustment
step 17, and cal.adjust.ac, Frequency step 18. The descriptions of these steps are described
in the following.
DC voltage adjustment steps 0 to 5
DC voltage adjustment step 0: MUX amp offset
Send the command:
cal.adjust.dc(0)
DC voltage adjustment step 1: Input 4-wire short circuit
1. Allow the instrument to settle for 30 seconds.
2. Send the command:
cal.adjust.dc(1)
DC voltage adjustment step 2: Open circuit
This step supports open terminals with attached cables. At the beginning of this step, the HI and LO
terminals are measured for applied voltage detection. If the voltage is ≤30 V DC and 1.0 V
1 KHz, the step will continue without a warning message
1. Remove the four-wire short from the inputs.
2. Send the following commands.
cal.adjust.dc(2)
DMM7510-905-01 Rev. A / April 2015
Figure 18: Front panel 4-wire shorting plug orientation
-- Supports either 4-wire short or open terminals.
-- Requires 4-wire short.
-- Requires open front terminals.
Section 3: Calibration
peak
3-11