Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
ACI Adjustment (cal.adjust.ac())
ACI Step 11, 100 µA 1 kHz, ACI 1 mA range zero
ACI Step 12, 1 mA 1 kHz, ACI 1 mA range full-scale and 10 mA range zero
ACI Step 13, 10 mA 1 kHz, ACI 10 mA range full-scale and 100 mA range zero
ACI Step 14, 100 mA 1 kHz, ACI 100 mA range full-scale, 1 A and 3 A range zero
ACI Step 15, 1 A 1 kHz, ACI 1 A range full-scale
ACI Step 16, 2 A 1 kHz, ACI 3 A range full-scale
AC-coupling amplitude adjustment
Frequency adjustment
ACV Step 18, 1 V 1 kHz, Squarewave, Frequency adjustment
For detail on how to use cal.adjust.ac(), refer to:
AC voltage adjustment steps 1 to 10
AC current adjustment steps 11 to 16
AC coupling adjustment step 17
AC frequency adjustment step 18
This command generates an error if:
Calibration is locked
The step is out of sequence
The step does not complete successfully
The value that is passed is invalid for the step, out of range, or not needed