Section 4: TSP command reference
Example
acal.schedule(acal.ACTION_RUN, acal.INTERVAL_1DAY, 8)
Sets auto calibration to run every day at 8 am.
Also see
Auto calibration
acal.run()
(on page 4-6)

cal.adjust.ac()

This function starts the specified AC adjustment step.
Type
TSP-Link accessible
Function
Yes
Usage
cal.adjust.ac(step)
cal.adjust.ac(step, value)
step
value
Details
The descriptions of these tests are:
ACV adjustment (cal.adjust.ac())
ACV Step 1, 10 mV 1 kHz, ACV 100 mV range zero
ACV Step 2, 100 mV 1 kHz, ACV 100 mV range full-scale and 1 V range zero
ACV Step 3, 100 mV 50 kHz, ACV 100 mV frequency flatness adjust
ACV Step 4, 1 V 1 kHz, ACV 1 V range full-scale and 10 V range zero
ACV Step 5, 1 V 50 kHz, ACV 1 V range frequency flatness adjust
ACV Step 6, 10 V 1 kHz, ACV 10 V range full-scale and 100 V range zero
ACV Step 7, 10 V 50 kHz, ACV 10 V range frequency flatness adjust
ACV Step 8, 100 V 1 kHz, ACV 100 V range full-scale and 700 V range zero
ACV Step 9, 100 V 50 kHz, ACV 100 V range frequency flatness adjust
ACV Step 10, 700 V 1 kHz, ACV 700 V range full-scale
4-8
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
(on page 2-4)
Affected by
The AC adjustment step to start
The value for this adjustment step (if applicable)
Where saved
Default value
DMM7510-905-01 Rev. A / April 2015