Keithley DMM7510 Calibration Manual page 56

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Verify DC current 10 µA and 100 µA range
Characterized shunt (Ω): 1.9607843E+05
Range
Ref DMM
Zero
1.00E-05
Ref DMM
Zero
1.00E-04
DMM7510-905-01 Rev. A / April 2015
Nominal
Calibrator
Shunt
input
setpoint
measure-
(V)
ment
(V)
0.00E+00
0
8.00E-07
0
STANDBY
n/a
0.00001
1.9707843
1.9607843
0.000005
0.9853921
0.9803922
-0.000005
-1.9707843 -0.9803922
-0.00001
-1.9707843 -1.9607843
0.00E+00
0 V
8.00E-07
0
0.0001
19.617843
19.607843
0.00005
9.8089216
9.8039216
-0.00005
-9.8089216 -9.8039216
-0.0001
-19.617843 -19.607843
Section 2: Performance verification
Calculated input
Lower limit
(A)
(A)
0
n/a
n/a
-3.00E-10
1.0000000E-05
9.9989500E-06
5.0000003E-06
4.9993253E-06
-5.0000003E-06
-5.0006753E-06
-1.0000000E-05
-1.0001050E-05
0
n/a
n/a
-9.00E-10
1.0000000E-04
9.9994000E-05
5.0000001E-05
4.9997001E-05
-5.0000001E-05
-5.0003001E-05
-1.0000000E-04
-1.0000600E-04
Upper limit
(A)
n/a
3.00E-10
1.0001050E-05
5.0006753E-06
-4.9993253E-06
-9.9989500E-06
n/a
9.00E-10
1.0000600E-04
5.0003001E-05
-4.9997001E-05
-9.9994000E-05
2-47