Section 2: Performance verification
Configure Digitize DC current for the 10 µA and 100 µA ranges:
1. Set up the Model DMM7510 for digitize DC current and the range being tested. Make sure
relative offset is disabled.
2. Verify the shunt. With a shunt characterization of 200 kΩ parallel 10 MΩ reference, the DMM
typical measurement is 196.07843 kΩ.
3. Connect the calibrator, shunt, Model DMM7510, and reference DMM as shown in the following
figure.
2-52
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Figure 10: Connections for the 10 µA and 100 µA shunt characterization
DMM7510-905-01 Rev. A / April 2015