Keithley DMM7510 Calibration Manual page 84

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Rear DC adjustment step 0: amplifier offset
Send the commands:
cal.adjust.rear.dc(0)
Rear DC adjustment step 1: 4-wire zero
Send the command:
cal.adjust.rear.dc(1)
Rear DC adjustment step 2: Open circuit
Remove the four-wire short from the inputs.
Cables cannot be attached to the Model DMM7510 during this step. At the beginning of this step, the
HI and LO terminals are measured for applied voltage detection. If detected, the step will terminate,
generating a warning messages. Also, any attached cables will significantly impact accuracy or
cause the adjust step to fail.
Send the command:
cal.adjust.rear.dc(2)
DMM7510-905-01 Rev. A / April 2015
-- Supports either 4-wire short or open terminals.
-- Requires 4-Wire short.
Figure 16: Rear-panel terminals open
-- Requires open rear terminals.
Section 3: Calibration
3-7