Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 2: Performance verification
Verify AC voltage accuracy for the 700 V range
Use shielded low capacitance cabling. For the 700 V range, avoid cable capacitances of >150 ρF.
Excessive capacitance may result in additional load regulation uncertainties and could cause the
calibrator output to open (go into standby).
To verify AC voltage accuracy for the 700 V range:
1. Put calibrator in Standby.
2. Connect the Model DMM7510 HI and LO INPUT connectors to the calibrator as shown in the
following figure.
3. For 700 V at 50 kHz and 100 kHz outputs, connect the calibrator to the Fluke 5725A amplifier.
DMM7510-905-01 Rev. A / April 2015
2-15