Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Verify digitize current 1 mA range
Description
UUT Zero
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
Verify digitize current 10 mA range
Description
UUT Zero
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
Verify digitize current 100 mA range
Description
UUT Zero
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
Verify digitize current 1 A range
Description
UUT Zero
Full scale (+)
Half scale (+)
Half scale (-)
Full scale (-)
Verify digitize current 3 A range
Description
UUT Zero
⅔ scale (+)
Half scale (+)
Half scale (-)
⅔ scale (-)
DMM7510-905-01 Rev. A / April 2015
Calibrator setpoint
Lower limit
(A)
STANDBY
-7.50E-08
0.001
9.99765E-04
0.0005
4.99845E-04
-0.0005
-5.00155E-04
-0.001
-1.00024E-03
Calibrator setpoint
Lower limit
(A)
STANDBY
-7.50E-07
0.01
9.99765E-03
0.005
4.99845E-03
-0.005
-5.00155E-03
-0.01
-1.00024E-02
Calibrator setpoint
Lower limit
(A)
STANDBY
-1.00E-05
0.1
9.99450E-02
0.05
4.99675E-02
-0.05
-5.00325E-02
-0.1
-1.00055E-01
Calibrator setpoint
Lower limit
(A)
STANDBY
-1.10E-04
1
9.99390E-01
0.5
4.99640E-01
-0.5
-5.00360E-01
-1
-1.00061E+00
Calibrator setpoint
Lower limit
(A)
STANDBY
-4.50E-04
2
1.99775E+00
1.5
1.49820E+00
-1.5
-1.50180E+00
-2
-2.00225E+00
Section 2: Performance verification
Upper limit
7.50E-08
1.00024E-03
5.00155E-04
-4.99845E-04
-9.99765E-04
Upper limit
7.50E-07
1.00024E-02
5.00155E-03
-4.99845E-03
-9.99765E-03
Upper limit
1.00E-05
1.00055E-01
5.00325E-02
-4.99675E-02
-9.99450E-02
Upper limit
1.10E-04
1.00061E+00
5.00360E-01
-4.99640E-01
-9.99390E-01
Upper limit
4.50E-04
2.00225E+00
1.50180E+00
-1.49820E+00
-1.99775E+00