Section 4: TSP command reference

cal.adjust.count

This attribute returns the number of times the instrument has been adjusted.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
adjustments = cal.adjust.count
adjustments
Details
You can use this command if calibration is locked or unlocked.
Example
Count = cal.adjust.count
print(Count)
Also see
cal.adjust.date
4-10
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Affected by
Not applicable
The number of adjustments
(on page 4-11)
Where saved
Nonvolatile memory
Assign the number of times the instrument has been adjusted
to a user variable named Count.
Output the value. Example output:
3
This shows that the instrument has been adjusted 3 times.
Default value
Not applicable
DMM7510-905-01 Rev. A / April 2015