Cal.password - Keithley DMM7510 Calibration Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual

cal.password

This attribute sets the password that you send when you unlock calibration.
Type
TSP-Link accessible
Attribute (W)
Yes
Usage
cal.password = password
password
Details
This command can only be sent when calibration is unlocked.
The password is not saved until calibration is saved with cal.save().
Be sure to record the password; there is no command to retrieve the password once it is set.
Example
cal.unlock("KI000CAL")
cal.password = "XYZCorp"
cal.lock()
cal.unlock("XYZCorp")
Also see
cal.save()
cal.unlock()
DMM7510-905-01 Rev. A / April 2015
Affected by
Not applicable
A string that represents the password to unlock calibration; must be 10 characters or
less
(on page 4-18)
(on page 4-19)
Section 4: TSP command reference
Where saved
Default value
Nonvolatile memory
KI000CAL
To change the default calibration password, unlock
the calibration with the default password.
Sets the password to XYZCorp.
Lock calibration. Use the password XYZCorp for
subsequent unlocks.