Cal.password - Keithley DMM6500 Manual

6 1/2-digit multimeter with scanning
Hide thumbs Also See for DMM6500:
Table of Contents

Advertisement

DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual

cal.password

This attribute sets the password that you send when you unlock calibration.
Type
TSP-Link accessible
Attribute (W)
Yes
Usage
cal.password = "password"
password
Details
This command can only be sent when calibration is unlocked.
The password is not saved until calibration is saved with cal.save().
Be sure to record the password; there is no command to retrieve the password once it is set.
Example
cal.unlock("KI000CAL")
cal.password = "XYZCorp"
cal.lock()
cal.unlock("XYZCorp")
Also see
cal.save()
(on page 4-8)
cal.unlock()
(on page 4-9)
DMM6500-905-01 Rev. E June 2021
Affected by
Not applicable
A string that contains the password to unlock calibration; maximum of 10 characters
Section 4: TSP command reference
Where saved
Default value
Nonvolatile memory
KI000CAL
To change the default calibration password, unlock
the calibration with the default password.
Sets the password to XYZCorp.
Lock calibration. Use the password XYZCorp for
subsequent unlocks.
4-7