Sterilisation; Handpiece Sterilisation; Tubing Sterilisation - KaVo HealOzone 2130 C Instructions For Use Manual

Dental ozone machine
Table of Contents

Advertisement

Instructions for use HealOzone 2130 C
6 Preparation methods DIN EN ISO 17664
CAUTION
CAUTION
|

6.3 Sterilisation

6.3 Sterilisation

6.3.1 Handpiece sterilisation

Damage caused by improper sterilisation.
Danger from contact corrosion.
▶ Remove product from steriliser immediately after the sterilisation cycle!
▶ Before sterilisation, detach handpiece from instrument tubing.
▶ Remove the endodontics cannula with the wrench from the handpiece.
▶ Sterilise the handpiece without the endodontics cannula.
▶ Always sterilise the endodontics cannula in the wrench.
▶ For the following procedures, sterilise the product in vapour-permeable paper or
material (Stericlav bag, KaVo).
Triple prevacuum
Gravity
Gravity
Drying time
▶ Spray the endodontics cannula an O-ring after sterilisation with KaVo spray
Mat. no. 0.411.9630.

6.3.2 Tubing sterilisation

Damage caused by improper sterilisation.
Danger from contact corrosion.
▶ Remove product from steriliser immediately after the sterilisation cycle!
▶ Before sterilisation, detach handpiece from instrument tubing.
▶ For the following procedures, sterilise the product in vapour-permeable paper or
material (Stericlav bag, KaVo).
Triple prevacuum
Gravity
Gravity
Drying time
at least 3 min
at least 10 min
at least 60 min
> 10 min
at least 3 min
at least 10 min
at least 60 min
> 10 min
44/57
134 ± 1ºC
134 ± 1ºC
121 ± 1ºC
134 ± 1ºC
134 ± 1ºC
121 ± 1ºC

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents