Instructions for use KaVo polymerisation light Poly One
6 Reprocessing methods according to ISO 17664 | 6.2 Disinfection
▶ Hold the curing light Poly One by the handpiece and disassemble the fibre-op‐
tic cable.
▶ Disassemble the silicone holder and protective shield.
▶ Clean the fibre-optic cable, protective shield, and silicone holder with a brush
and running water (at least tap water quality).
▶ Immediately remove contamination from materials used in the dental office us‐
ing a soft cloth and alcohol.
▶ Clean the handpiece with a soft, moist cloth.
6.1.2 Automated cleaning
CAUTION
Damage caused by improper cleaning/disinfection.
Do not sterilise the fibre-optic cable in a hot-air steriliser.
▶ Subject the fibre-optic cable to thermodisinfection and sterilisation.
6.2 Disinfection
6.2.1 Manual disinfection
External disinfection
Approved disinfectants:
▪ Mikrozid AF Liquid (made by Schülke & Mayr)
▪ FD 322 (Dürr)
▪ Incidin Liquid (Ecolab)
▪ CaviCide (Metrex)
18/22